|
Traceability & Uncertainty
of Measurement
Below is a summary of
L-A-B Policy 001 which defines the way in which laboratories will prove
their Traceability of Measurements, and express their Measurement
Uncertainty. Please refer to
L-A-B Policy 001 for the exact requirements of Laboratory Accreditation
Bureau.
Traceability
-
The calibration program shall assure
traceability of measurements, and/or verification and validation of
equipment is traceable, wherever possible, to NIST or other National
Measurement Institute (NMI) is to be captured within L-A-B Form 001 -
Traceability Tracking.
Calibration certificates, where applicable, shall indicate the
traceability to an NMI or intrinsic standard, the measurement result and
the associated uncertainty of the measurement and/or a statement of
compliance with an identified metrological specification. To ensure
actual traceability, the path of reference standard verification back to
the NMI shall be clear. The evidence of the investigation of the path
back to the NMI shall be available for verification by the L-A-B
assessors. The requirements for a primary reference, transfer, and
working standards or reference materials shall be defined by the
laboratory. When defining those requirements the laboratory shall
identify the critical characteristics that may affect the traceability
for the calibration and/or test. Those characteristics include the
requirements stated in ISO/IEC 17025. Critical characteristics may
include handling, reporting, equipment, and methodology using the
standards etc. Depending on the level of standard and the frequency of
use, transport, ownership, etc. the laboratory shall apply the
appropriate degree of procedural control.
Where intrinsic standards are used, the laboratory should demonstrate by
measurement-assurance techniques, interlaboratory comparison, or other
suitable means that its intrinsic-measurement results are correlated
with an NMI.
-
When traceability to an
NMI is not possible, the laboratory should have a procedure that will
provide satisfactory evidence that the results are correlated, for
example by participation in a suitable interlaboratory comparison or
proficiency testing. Other satisfactory evidence would be an
internationally accepted standard in the field concerned; suitable
reference material; ratio or reciprocity-type measurements; or mutual
consent standards that are clearly specified and mutually agreed upon by
all parties concerned.
Demonstration of Traceability
-
L-A-B prospective clients and accredited
labs may submit appropriate physical standards and test and measurement
equipment (M&TE) directly to NIST or when appropriate, to another
national metrology institute (NMI). Accredited laboratories may obtain
certified reference materials from NIST (called Standard Reference
Materials (SRM) under copyright) or from another NMI. Use of an NMI
other than NIST must be documented as the appropriate NMI relevant for
the scope of accreditation and stated uncertainties.
-
Testing laboratories that perform
calibrations only for themselves do not need to be accredited as a
calibration laboratory. Calibration laboratories that perform specific
calibrations only for themselves to support their accredited services do
not need to be accredited for those calibrations. For the purpose of
assuring traceability, an accredited laboratory may calibrate its own
equipment if the appropriate requirements of L-A-B and 17025 have been
met. The laboratory shall demonstrate is competency to perform the
calibrations it undertakes.
-
L-A-B prospective clients and accredited
labs may use a State Weights and Measure lab that is recognized by the
NIST Office of Weights and Measures (OWM) State Laboratory Program and
has a current “Certificate of Measurement Traceability” issued by the
Weights and Measures Division of NIST. This certificate must be
available during the assessment of the laboratory.
-
L-A-B Accredited Laboratories that do not
demonstrate traceability as defined in 1, 2 or 3 above, shall use
accredited calibration laboratory services wherever available.
Accredited calibration laboratories are those accredited by L-A-B or an
accreditation body that is recognized as a signatory of the Asia Pacific
Laboratory Accreditation Cooperation “APLAC” MRA and/or International
Laboratory Accreditation Cooperation “ILAC” MRA. A listing of L-A-B
accredited laboratories is available on our website at www.l-a-b.com.
When utilizing accredited calibration laboratory services, the
calibration certificates shall be accompanied by a recognized
accreditation body symbol or otherwise make reference to accredited
status to be considered satisfactory for traceability purposes.
-
If a L-A-B
applicant or accredited laboratory submits physical standards or M&TE to
a calibration provider that is not accredited by L-A-B or other L-A-B
acceptable laboratories, the laboratory shall:
-
Document
that an appropriate accredited calibration provider is not
available. Documented evidence is required including results of
searches of appropriate AB’s websites.
-
Audit
the claim of traceability of the provider of the calibration service
and document the following areas related to the calibration and
claim of traceability of its standards and M&TE:
-
Information
regarding assessment of the quality system used by the
calibration service provider. The information must include
details on an assessment of the calibration service provider.
The assessment can be done by the laboratory (if a qualified
person is on staff) or another AB.
-
The
calibration procedure(s) used by the calibration service
provider.
-
The
physical standards or other M&TE used by the calibration service
provider (including evidence of traceability to standards
maintained by NIST or an appropriate NMI and copies of relevant
calibration certificates).
-
Information regarding the calibration intervals of relevant
standards or other M&TE.
-
The
environmental conditions of the laboratory.
-
The
method(s) by which uncertainties are determined e.g., ISO Guide
to the Expression of Uncertainty in Measurement (GUM).
-
The
relative uncertainties achieved at all steps of the process.
-
Pursue
the traceability chain until traceability to appropriate stated
references is completely validate, when a calibration service
provider submits physical standards and/or M&TE used in the
calibration to another laboratory(s) not accredited by L-A-B or
L-A-B acceptable accreditation body.
-
Enter
the audit documentation, including all findings of noncompliance and
resolutions of those findings, into the laboratory’s quality
management record system.
-
If
traceable calibration is not available or appropriate, laboratories
may demonstrate comparison to a widely used standard that is clearly
specified and mutually agreeable to all parties concerned,
particularly in measurements where NIST does not maintain a US
national standard. For example, NIST does not maintain a standard
for all hardness testing scales. There are several widely used
commercial standards available for hardness. However, these
standards may not all give equivalent measurement results;
therefore, it is important to specify which standard is used and to
obtain agreement among all parties involved that the choice made is
acceptable
Measurement Uncertainty
L-A-B has created a new
uncertainty Budget Checksheet to help you through the steps of creating an
uncertainty budget.
-
Uncertainty
of measurement comprises, in general, many components. Some of these
components may be estimated on the basis of the statistical distribution
of the results of series of measurements and can be characterized by
experimental standard deviations, and are called Type A evaluation. The
Type A can be applied when several independent observations have been
made for one of the input quantities under the same conditions of
measurement. If there is sufficient resolution in the measurement
process, there will be an observable scatter or spread in the values
obtained. In this case, the standard uncertainty is the experimental
Standard Deviation of the mean that follows from an averaging procedure
or an appropriate regression analysis.
-
Estimates of
other components can only be based on experience or by scientific
judgment based on all available information on the possible variability
of the measurement and other information, and are called
Type B evaluation. Values belonging in this category may be derived
from:
-
Previous
measurement data.
-
Experience with or general knowledge of the behavior and properties
of relevant materials and instruments.
-
Manufacturer’s specifications.
-
Data
provided in calibration and other certificates.
-
Uncertainties assigned to reference data taken from handbooks.
3.
The measurement uncertainty for a given calibration is the combination of
all the Type A and Type B components
of the uncertainty budget.
Best
Measurement Capability
Best
measurement capability
(always referring to a particular quantity, viz. the measurand) is
defined as the smallest uncertainty of measurement that a laboratory can
achieve within its scope of accreditation, when performing more or less
routine calibrations of nearly ideal measurement standards intended to
define, realize, conserve or reproduce a unit of that quantity or one or
more of its values, or when performing more or less routine calibrations of
nearly ideal measuring instruments designed for the measurement of that
quantity. The assessment of best measurement capability of accredited
calibration laboratories is based on the method described in this document,
and shall be supported or confirmed by experimental evidence.
-
L-A-B
requires that calibration accredited laboratories state on their Scope
of Accreditation an expanded uncertainty of measurement U, obtained by
multiplying the standard uncertainty u(y) of the output estimate by a
coverage factor of k,
U=ku(y)
In cases where a normal (Gaussian) distribution can be attributed to the
measurand and the standard uncertainty associated with the output
estimate has sufficient reliability, the standard coverage factor k=2
shall be used. The assigned expanded uncertainty corresponds to a
coverage probability of approximately 95%. These conditions are
fulfilled in a majority of cases encountered in calibration work.
Calibration
and Dimensional Inspection Laboratories
-
Calibration
and dimensional
inspection laboratories shall report their measurement uncertainty on
all calibration certificates, and inspections reports unless it can be
proven that the client does not want it reported. Evidence that the
client does not want the calibration and dimensional inspection
uncertainty reported shall be available for an assessor to review at the
time of an assessment. Regardless of whether the client wants the
measurement uncertainty reported, the laboratory shall retain sufficient
information to report the uncertainty.
-
Laboratories may issue certificates with a
statement of compliance (or conformance to a specification). When the
laboratory issues a statement of compliance it must ensure that:
-
The specification is a national /
international standard or one that has been provided or agreed to by
the customer.
-
The measurements needed to determine
conformance are within the scope of accreditation.
-
When the measurement result is
determined to be within a specified tolerance, the associated
uncertainty of the measurement result is properly taken into account
with respect to the tolerance by a documented procedure or policy.
The policy or procedure must be established and implemented by the
laboratory that defines the decision rules used by the laboratory
for declaring in or out of tolerance conditions.
-
The default decision rule is found
in ILAC-G8:1996, Guidelines on Assessment and Reporting of
Compliance with Specification, section 2.5. With agreement from
the customer, other decision rules may be used as provided for
in this section of the Requirements.
-
The certificate relates only to
metrological quantities and states which clauses of the
specification are certified to have been met.
-
The calibration laboratory shall report
its best measurement capability, as defined above, on its proposed Scope
of Accreditation.
-
L-A-B does allow the best measurement
capability to be reported on the appropriate calibration/dimensional
inspection certificate only if the best measurement capability was
computed (and reported on the scope) at the fullest extent of the
measuring range representing the worst capability of the lab to perform
that measurement. Other capabilities (uncertainties) that may be worse
than the one reported on the scope of accreditation shall be reported on
the certificates of calibration or dimensional inspection if they more
appropriately represent the measurement taken.
Testing
Laboratories
-
The
complexity involved in estimation of uncertainty of measurement in the
case of testing varies considerably from one test field to another and
also within one field itself. A less metrologically rigorous process
than that which can be followed for calibration can also often be used.
Clause 5.4.6.2 of ISO/IEC 17025 allows for these factors. The degree of
rigor needed in an estimation of uncertainty of measurement depends on
factors such as:
-
Requirement of the test method
-
Requirement of the client
-
There
are narrow limits on which decisions conformance to a specification
are based
-
If the test
method is well recognized (ASTM, ISO) and specifies limits to the values
of the major sources of uncertainty of measurement, and specifies the
form of presentation of calculated results, the laboratory is considered
to have satisfied this clause by following the test method and reporting
instructions.
-
Uncertainty
components/budgets are a combination of many factors that may include,
but are not limited to:
-
Reference standards
-
Reference materials
-
T/C
methods used
-
Equipment used
-
Environmental conditions,
-
Properties and condition of item being tested
-
Calibration
-
Operator
-
Known
physical characteristics of components such as, coefficient of
thermal expansion. These often can be looked up in engineering and
scientific handbooks.
Requirements for Reporting Measurement Uncertainty in Testing Labs
-
The
laboratory shall perform and have available for the assessor a Needs
Assessment, as defined below. They shall also produce a procedure(s) for
calculating their measurement uncertainty, which includes creation of an
uncertainty budget for the assessor to review during your assessment or
surveillance visit.
-
All
necessary measurement uncertainties, as defined in the Needs Assessment,
shall be available for review by the assessor during the assessment or
surveillance visit. In the event that the calculations and uncertainty
budgets are not available, the laboratory runs the risk of having those
tests, for which the uncertainty is missing, removed from their Scope of
Accreditation.
-
Laboratories applying for accreditation during the above periods shall
have the appropriate documentation that relates to the dates defined
above.
-
Currently
L-A-B does not require reporting of measurement uncertainty on Test
Reports, unless it is required/requested by the laboratory’s client.
Needs Assessment
The laboratory
shall do a Needs Assessment for all tests on their proposed Scope of
Accreditation in the “Remarks” column for each “Field of Testing”. The Needs
Assessment shall define what actions will be necessary with regard to the
reporting of uncertainty. The Needs Assessment matrix shall group the tests
into five categories as defined below.
-
Qualitative
or semi-quantitative tests that require no uncertainty budgets
-
A test
performed to well-recognized test methods that specify limits to the
values of the major sources of uncertainty of measurement and specifies
the form of presentation of calculated results. These are defined in
ISO/IEC 17025 Clause 5.4.6.2 Note 2.
-
Chemical,
environmental, or biological test methods based on published regulatory
or consensus methods; such as, FDA, EPA, AOAC, and ASTM, for which
uncertainty is not defined in the method. For these types of test,
uncertainty can be estimated using the standard deviation of laboratory
control samples for more than 50 points. (This does not include
laboratory-developed methods that require validation and are covered
below).
-
Test methods
that need identification of the major components of uncertainty and a
reasonable estimate of the measurement uncertainty.
-
Test methods
that need identification of all components of uncertainty and detailed
measurement uncertainty budgets calculated in accordance with published
methods that are consistent with the ISO “ Guide to the Expression of
Uncertainty of Measurements”
|